Research ArticleBioengineering and Diabetes

Function of an Implanted Tissue Glucose Sensor for More than 1 Year in Animals

See allHide authors and affiliations

Science Translational Medicine  28 Jul 2010:
Vol. 2, Issue 42, pp. 42ra53
DOI: 10.1126/scitranslmed.3001148

Abstract

An implantable sensor capable of long-term monitoring of tissue glucose concentrations by wireless telemetry has been developed for eventual application in people with diabetes. The sensor telemetry system functioned continuously while implanted in subcutaneous tissues of two pigs for a total of 222 and 520 days, respectively, with each animal in both nondiabetic and diabetic states. The sensor detects glucose via an enzyme electrode that is based on differential electrochemical oxygen detection, which reduces the sensitivity of the sensor to encapsulation by the body, variations in local microvascular perfusion, limited availability of tissue oxygen, and inactivation of the enzymes. After an initial 2-week stabilization period, the implanted sensors maintained stability of calibration for extended periods. The lag between blood and tissue glucose concentrations was 11.8 ± 5.7 and 6.5 ± 13.3 minutes (mean ± standard deviation), respectively, for rising and falling blood glucose challenges. The lag resulted mainly from glucose mass transfer in the tissues, rather than the intrinsic response of the sensor, and showed no systematic change over implant test periods. These results represent a milestone in the translation of the sensor system to human applications.

Introduction

Diabetes is a major health problem that results in significant mortality, debilitating complications, substantial economic impact on society, and untold waste of human resources (1). The results of the prospective Diabetes Control and Complications Trial show that complications of diabetes can be significantly reduced by improved blood glucose control (2). Achieving improved glucose control is problematic for most people with the disease, however, because the most common means for measurement of blood glucose involves blood collection by “fingersticking,” a method that is inconvenient and unacceptable to many people with diabetes, and is rarely performed frequently enough to follow blood glucose dynamics.

Continuous glucose monitoring can now be performed with short-term, percutaneous glucose sensors (3), but this method has certain drawbacks. These sensors are inserted in subcutaneous tissues with needle introducers and remain in place for 3 to 7 days before being replaced. Their performance can be affected by changes that occur in the tissues as a result of the insertion and presence of the implant (4), which can lead to instability of the glucose signal, and fingerstick glucose assays are needed for regular sensor recalibration during use. These sensors have therefore not been approved by the Food and Drug Administration as a primary standard for glucose measurement and cannot be fully relied upon to warn the user of an impending hypoglycemic episode (5). Nevertheless, an important result of the use of short-term subcutaneous glucose sensors is the realization that blood glucose varies to a much greater extent during the course of the day than was previously thought, even in presumed well-controlled individuals. These sensors have also made possible initial clinical research on the artificial pancreas (6).

For acceptance by a broader group of users, there is a need for a long-term, totally implanted glucose sensor and wireless telemetry system. The implant should monitor tissue glucose continuously and report via telemetry to an external receiver that could display the blood glucose information or relay it to a caregiver. This information could be used in a number of ways to help achieve improved blood glucose control: It could direct dosing of therapeutics, warn of hypoglycemia, guide diet modification and exercise, or act as an input to an artificial pancreas. It could also be used in conjunction with other forms of therapy such as drugs, transplants, islet replacement or preservation, or stem cells. To be optimally acceptable by users, the sensor should function for a year or more, be insertable by a simple outpatient procedure requiring only local anesthesia, be convenient and unobtrusive, be free of significant risk for untoward effects, and not require frequent recalibration.

We report long-term tissue glucose monitoring with a sensor telemetry system implanted in subcutaneous tissues of pigs. Monitoring was carried out while pigs were nondiabetic (for 3 weeks in one animal and nearly 1 year in the other) and, after the pigs had been converted to insulin-dependent diabetic by administration of streptozotocin, continued in each animal for an additional 6 months, with diabetes managed by insulin injection and diet control.

Results

Sensing principle

The glucose sensor is based on the following two-step chemical reaction catalyzed by glucose oxidase and catalase (7):glucose+O2+H2Ogluconic acid+H2O2H2O212O2+H2O resulting in the overall enzyme reaction:glucose+12O2gluconic acid

The two enzymes are immobilized within a cross-linked protein gel that is in contact with an electrochemical oxygen sensor. Glucose and ambient oxygen diffuse into the gel and encounter the enzymes, the above reactions occur, and oxygen that is not consumed in the process is detected by the sensor. After comparison with the background oxygen concentration detected by a separate oxygen reference sensor, the difference is related to glucose concentration. The sensing unit is therefore composed of (i) a glucose sensor, which is an oxygen sensor with an immobilized enzyme gel that produces a glucose-modulated, oxygen-dependent current (igmo); (ii) a reference oxygen sensor without enzymes that produces an oxygen-dependent current (io); and (iii) a signal-processing element that takes the difference of (i) and (ii) to give the signal of interest—the glucose-dependent difference current (ig).

The sensor is calibrated with the formula BG = k1ioF(k2ig/io), where BG is blood glucose, k1 is the oxygen mass transfer coefficient for the reference sensor, k2 is the glucose mass transfer coefficient related to the implant environment, and F is a quasi-linear sensitivity function of the glucose sensor determined in vitro. With an appropriately designed immobilized enzyme gel structure (8), the sensor can remain responsive to glucose in the tissue implant environment over clinically relevant concentration ranges. This calibration relation is an adaptation of our previous methods (9, 10).

The implant

Eight 300-μm platinum working electrodes, with associated platinum counterelectrodes and Ag/AgCl potential reference electrodes, are arranged as four sensor pairs on the surface of a 1.2-cm-diameter alumina disc (11). The electrodes are covered by a thin electrolyte layer, a protective layer of medical-grade polydimethylsiloxane (PDMS), and a membrane of PDMS with wells for the immobilized enzymes located over certain electrodes. The enzymes are immobilized in the wells by cross-linking with albumin using glutaraldehyde, and the resulting gel is rinsed extensively to remove unbound material. The alumina disc is fused into a hermetically sealed titanium housing (Fig. 1) containing a potentiostat and signal-conditioning circuitry for each sensor, a wireless telemetry system, and a battery having a minimum 1-year lifetime. The implant is sterilized with a chemical sterilant by a procedure that has been validated according to standard methods (12).

Fig. 1

Sensor array with integrated telemetry system before implantation. The implant is 3.4 cm in diameter and 1.5 cm thick. The top surface of the implant includes two polyester velour patches for tissue adhesion. Cross-sectional schematic view shows electronics modules (A), telemetry transmission portal (B), battery (C), and sensor array (D).

The telemetry system samples the currents from individual sensors, encodes the samples into multiplexed signal segments, and transmits the segments as a train of radio-frequency signals at regular 2-min intervals to an external receiver, where the signals are decoded and recorded. The potentiostat circuitry in the implant is based on an electronic design described in detail elsewhere (13), extended in this case for control of eight individual working electrodes. Radiotelemetry is accomplished with a 433.92-MHz carrier signal, with a total effective radiated power of <100 nW, providing a practical effective transmitter range of ~10 feet (3.085 m) with information packet reception rate exceeding 97%. This hermetically sealed wireless telemetry system makes possible long-term recordings without infection-prone percutaneous electrical leads.

Implantation

Two series of implant studies were conducted. In the first series, intended to aid design optimization and component reliability verification, a total of 30 individual sensor telemetry units were implanted in six nondiabetic pigs to refine the surgical technique, evaluate device tolerance and biocompatibility, test the electronic circuitry and telemetry, and identify factors that affect the lifetime of the sensor. The devices in this series were explanted and analyzed according to preset protocol schedules at periods ranging from 1 to 18 months after implantation. Results from this foundational research included verification of (i) acceptable long-term biocompatibility, assessed after 18-month implant periods; (ii) immobilized enzyme life exceeding 1 year; (iii) battery life exceeding 1 year; (iv) electronic circuitry reliability and telemetry performance; (v) sensor mechanical robustness including long-term maintenance of hermeticity; (vi) stability of the electrochemical detector structure; and (vii) acceptability and tolerance of the animals to the implanted device. Data from this study are available from the authors upon request. Results were also obtained from this series related to the effects of tissue permeability and tissue remodeling and are reported below.

In the second series, which involved evaluation in diabetic conditions and which is described in detail here, two devices were implanted in each of two pigs (four devices total) and first operated for 352 days (subject 1) and 16 days (subject 2), respectively, with the animals in the nondiabetic state. The animals were then made diabetic by administration of streptozotocin, and the devices continued to operate for an additional 168 days in subject 1 to a total of 520 days, and for an additional 206 days in subject 2 to a total of 222 days. Individual experiments were terminated because of the substantial resources required to maintain diabetic animals.

No adverse medical events (infection, erosion, migration, etc.) were encountered with any implant in either test series. Together, the test series represent a collective 31 total device-years of implant experience, with 17 of the devices remaining implanted and functional for more than 1 year.

Glucose sensor recordings

The sensor response in subject 1 over a 5-week period during the nondiabetic phase is shown in Fig. 2. In this example, only one system calibration adjustment was performed during the period (on day 186), with the goal of obtaining a qualitative indication of the implanted sensor stability. From that point forward, a fixed calibration regimen with regular calibration adjustment every 10 days was used for analysis of sensor accuracy (see below). A practical recalibration schedule will need to be established later for human use.

Fig. 2

Long-term continuous monitoring in nondiabetic and diabetic pigs. Timelines for sensor operation in subjects 1 and 2 are provided at the top; diabetes induction is indicated for each animal by an arrow. Sensor outputs are shown as solid red lines, and plasma glucose values from laboratory analysis of central venous samples are shown as solid blue circles. Left: The displayed 5-week period begins 23 weeks after implantation in subject 1. Plasma glucose values were sampled during intravascular glucose tolerance tests (IVGTTs) administered once or twice weekly to assess the sensitivity to glucose during the nondiabetic phase. One system calibration adjustment was performed during the entire period (on day 186). Glucose concentrations are relatively stable between IVGTTs in nondiabetic pigs despite eating, and intravenous glucose challenges were required to produce significant glucose excursions. Right: The sensor output is shown over two continuous 3-day periods in two diabetic pigs. In subject 1, the sensor had been operated for 352 days with the animal in the nondiabetic state at the time of induction of diabetes, after which monitoring continued for 168 more days. The segment displayed begins on day 373 after sensor implantation (21 days after conversion of the animal to diabetic). In subject 2, diabetes was induced 16 days after device implantation, after which monitoring continued for another 206 days. The segment displayed begins on day 19 after sensor implantation (3 days after conversion of the animal to diabetic).

Results after the pigs became diabetic are also shown in Fig. 2. As expected, we noted a considerable difference in the extent and duration of glucose excursions between the nondiabetic and the diabetic states. In the nondiabetic state, glucose concentration remained relatively constant at a baseline of ~75 mg/dl, regardless of feeding and physical activity. As a result, to test the sensor sensitivity to glucose, it was necessary to create blood glucose excursions by intravenous infusion of glucose, resulting in rapid rise and rapid unaided return to the baseline caused by the endogenous insulin response. In the diabetic state, however, the blood glucose concentration varied substantially with time, rising and falling in complex response to feeding, activity, and administration of insulin. (Note that the sensor response was electronically “capped” at 400 mg/dl, so glucose values above that level are not reported by the sensor.) Insulin injections were needed regularly to interrupt sustained hyperglycemia episodes.

Accuracy

We assessed the sensor accuracy on the basis of data collected from glucose excursion tests conducted during the diabetic phase in subjects 1 and 2. Conventional statistical methods were used, including standard regression analysis (14), error grid plots that segregate results into graphical regions on the basis of potential clinical significance by using both original sensor values and values adjusted for delay (described below) (15), and mean and median absolute relative difference (ARD) analyses (16). Results obtained from the diabetic animals are as follows (with values retrospectively adjusted for an average 6.6-min delay, determined for the diabetic phase as discussed below, in parentheses): number of points: 392; error grid values: 63.8% (70.4%) of points in region A (error has no effect on clinical action), 32.4% (28.6%) in region B (error is clinically benign), 3.6% (1.0%) in region C (error likely to affect clinical outcome), 0.3% (0%) in region D (error poses medical risk), and 0% (0%) in region E (error is potentially dangerous clinically); ARD values: mean, 22.1% (17.9%); median, 14.7% (13.2%); and correlation coefficient: 0.88 (0.92). These results suggest that, although there may be quantifiable differences between the actual blood glucose concentrations and values reported by the sensor, none of these differences would lead to a mistaken, potentially dangerous clinical action. The results obtained here are comparable to published values obtained for shorter periods from currently available short-term continuous glucose monitors used clinically (16, 17).

Signal delay

When sensors are operated continuously, a lag or delay exists during dynamic conditions between the actual blood glucose concentration and the value reported by the sensor. Associated with the delay is the dynamic error, which is the difference between the actual and reported blood glucose values at a given time (18). The delay and associated error can depend on the intrinsic rate of the sensor response, the rate of glucose mass transfer in the tissues, and the instantaneous rate of blood glucose change of the subject (18). The component of delay attributable to the sensor is determined from in vitro experiments in which glucose concentration is changed abruptly. The instantaneous rate of blood glucose change is determined by frequent blood glucose sampling, and the rate of glucose mass transfer in the tissues is limiting when the other two processes are much faster.

The delay in the response to an intravascular glucose tolerance test (IVGTT) in subject 1 on day 168 during the nondiabetic phase is shown in Fig. 3. The maximal rate of glucose rise because of glucose infusion was ~8 mg/dl per minute, and the maximal rate of fall because of endogenous insulin action was 6 mg/dl per minute. After an initial lag, the sensor signal rose at a rate parallel to the plasma glucose ramp. The glucose concentration then remained at a plateau value of ~260 mg/dl created by a reduced glucose infusion rate for ~15 min, before falling toward the baseline, with the sensor signal falling thereafter. During the excursion tests in the diabetic phases, maximal rates of central venous plasma glucose change (mean ± SD, n = 34) were 4.1 ± 1.9 and 5.2 ± 1.0 mg/dl per minute for rising and falling transitions, respectively. The rates of blood glucose change during testing in the nondiabetic phases were significantly more rapid than the maximal spontaneous rates of change reported in diabetic subjects (19), which are ~3 mg/dl per minute rising and 2.5 mg/dl per minute falling.

Fig. 3

Example of sensor response during an IVGTT excursion. Plasma glucose values in solid blue circles are connected by the blue line and the sensor signal is the red line. The arrows indicate the delay at the 50% points between the minimum and the maximum plasma glucose concentrations for the rising and falling excursions.

The definition of the rising or falling delay used here (shown by the arrows in Fig. 3) is the time between the plasma glucose value and the sensor value at the 50% point between the minimum and maximum plasma glucose values in an excursion. For example, if the minimum plasma glucose value in an excursion was 100 mg/dl (for example, at baseline, before infusion of glucose) and reached a plateau at 200 mg/dl upon glucose infusion, the rising delay is the difference between the time the plasma glucose reaches 150 mg/dl and the time that the sensor indicates 150 mg/dl. The falling delay from each excursion was evaluated during the falling leg at the same 50% plasma glucose crossing point. With this technique, the average value of the rising delay was 11.8 ± 5.7 min (mean ± SD) and of the falling delay was 6.5 ± 13.3 min, based on 34 IVGTTs in subject 1 during the nondiabetic period. Of these values, 2.5 ± 1.2 min is ascribable to the sensor itself, as determined from independent in vitro measurements, and an estimated 0.5 min is ascribable to circulatory transport from the central venous infusion site to the implant site. The remainder of the rising and falling average delays (8.8 and 3.5 min, respectively) is attributable to mass transfer and physiologic phenomena within the local tissues. Over the extended implant period, there was no change in either average delay value.

We confirmed these delay values with an alternative approach (20) based on systematic retrospective displacement of the sensor signal values with respect to the measured plasma glucose values and determination of the root mean square coefficient of variation between all sensor and plasma values at each step. For the IVGTT data set of subject 1 referenced above, the minimum coefficient of variation value was obtained at a signal displacement of ~10 min, which is comparable to the average of the rising and falling lag values determined as reported above. In the diabetic phase, the minimum coefficient of variation value was obtained at a signal displacement of 6.6 min (average of subjects 1 and 2).

Oxygen reference sensor

Signals from the oxygen reference sensor indicate the time course of change in tissue permeability after implantation. The averaged signals from the oxygen reference sensors from the series of implanted animals are plotted as a function of implant time in weeks (Fig. 4). Each data point (open circle) represents an average of 60 sensor signals at the indicated time after implantation, and points are fitted to an exponential decay curve (black line). It is noteworthy that the averaged oxygen signals decay exponentially, asymptotically approaching a nonzero value within ~6 weeks, and remain relatively constant thereafter. Previous studies with hamsters (21) showed that the exponential signal decay is due to changes in the effective permeability of the tissue, rather than changes in the sensitivity of the sensors per se. This was demonstrated (21) by comparison of preimplantation and postexplantation measurements of sensor sensitivity to oxygen in the gas phase, where boundary layers are absent and highly precise measurements are possible. Thus, the decay of both the oxygen reference signal and the oxygen component of the glucose sensor signal is due to changes in the effective permeability of the tissue, which stabilizes within several weeks.

Fig. 4

Signals from implanted oxygen sensors over 3 months. The averaged signal (open circles), expressed as oxygen electrode current in nanoamperes, which is proportional to the permeability of the foreign body tissue to oxygen, decays exponentially, approaching an asymptotic constant value at ~6 weeks. The exponential fit (black line) is provided by io = a exp(−bt) + c, where a = 4.5 nA, b = 0.49 week−1, c = 1.0 nA, and R2 = 0.96. Each data point is an average of the signals from 60 electrodes. Data comprising the 10th and 90th percentiles are represented by the yellow and red lines, respectively.

The observation that the sensitivity to glucose remains stable during a period of significant oxygen signal decay reveals an advantage of our sensor design. The glucose-sensing strategy used here, which is based on differential oxygen detection, reduces the sensitivity of the glucose-dependent signal to tissue encapsulation by the foreign body response to the implant. Because the substrate sensitivity of both sensors decays in parallel as the effective tissue permeability decreases after implantation, the glucose-dependent difference signal remains largely unaffected.

Tissue remodeling

Five-millimeter-thick tissue samples were collected from two pigs at the time of implant removal on day 91, fixed by using standard methods (see Materials and Methods), and stained with hematoxylin and eosin (H&E) and Gomori trichrome stains and antibodies to visualize nuclear density, collagen density, and microvascular density (Fig. 5). Medical-grade titanium, a well-known biocompatible material (22) that we used for the implant housing, served as a control. Nuclear density and collagen density are shown for tissues adjacent to the PDMS membrane surface (Fig. 5, upper panels) and titanium surface (Fig. 5, lower panels). The quantitative results (mean ± SD, n = 5) were as follows: nuclear (or cell) density, 1282 ± 201 nuclei/mm2 for tissue adjacent to the PDMS surface and 2769 ± 491 nuclei/mm2 adjacent to the titanium control; collagen density, 95.9 ± 2.1% (area ratio) adjacent to the PDMS surface and 53.1 ± 7.7% adjacent to the titanium control; microvascular density, 0.05 ± 0.08% (area ratio) adjacent to the PDMS surface and 0.04 ± 0.06% adjacent to the titanium surface. The lower nuclear density and higher collagen density adjacent to the PDMS surface indicate a greater degree of tissue remodeling adjacent to the PDMS material compared to the titanium surface. However, microvascular density was comparable for the two materials, and microvessels necessary for mass transfer were present within a short distance from the implant surface.

Fig. 5

Histology of tissue adjacent to the PDMS (upper panels) and titanium (Ti) (lower panels) surfaces. The interface with the implant is above. (a) H&E stain. Nuclear features are dark blue/purple, muscle and extracellular spaces are purple/pink, and adipocytes are white. (b) Gomori trichrome stain. Collagen deposits are blue/green, and nuclei are dark blue/purple. Red blood cells appear red. (c) CD31 stain. CD31 antibodies are dark brown/black, and nuclei are blue. Hematoxylin counterstaining of endothelial nuclei appears blue/purple.

Discussion

Design for long-term function

This sensor has several important design features that make possible long-term operation in the tissue environment. First, glucose oxidase is specific for glucose over other biochemicals present in tissue fluids, and with proper design, the sensor can therefore also be specific. Second, there is a nonporous PDMS layer between the enzyme membrane and the electrodes, which allows oxygen passage by solubilization in its hydrophobic phase but prevents electrode poisoning and interference from polar endogenous biochemicals and common exogenous chemicals such as acetaminophen and ascorbic acid. Third, the electrochemical oxygen sensors are based on the three-electrode potentiostatic principle (23) and retain long-term stability of oxygen sensitivity, in contrast to some conventional oxygen sensor systems. Fourth, glucose oxidase is inactivated by hydrogen peroxide, the catalytic product (24), but we have extended the lifetime of immobilized glucose oxidase by including coimmobilized catalase in excess to prevent peroxide-mediated inactivation (25) and by incorporating a large reserve of the enzymes to maintain a diffusion-limited design (26). These features are not feasible in other glucose sensor designs based on electrochemical detection of hydrogen peroxide (27).

Minimization of tissue irritation

The acceptable level of tissue remodeling may also be due, in part, to several design features of the sensor. The sensor membrane materials are biocompatible by standard in vitro biocompatibility tests (28) and release few, if any, irritants into the tissue. The pore-free PDMS layer prevents passage of current from the electrodes into the tissues and eliminates possible exacerbation of tissue encapsulation due to electrical flux, which may be a problem for some other implanted sensors. In the present design, catalase consumes peroxide, which would otherwise diffuse into adjacent tissues and cause strong irritation. Inclusion of catalase is not possible in other enzyme electrode sensors that are based on hydrogen peroxide detection (27). The implantation method also emphasizes blunt dissection, which is intended to avoid bleeding and minimize damage to the vasculature and lymphatic drains.

Oxygen access

The stoichiometric shortage of oxygen in tissues with respect to glucose, known as the oxygen deficit (29), can be two or more orders of magnitude. If not resolved, this discrepancy would cause the enzyme reaction in the sensor to be limited by oxygen rather than glucose, and the range of sensitivity to glucose to be substantially reduced. Our sensor design avoids this problem both by salvaging one-half equivalent of oxygen from hydrogen peroxide via the catalase reaction and by controlling the relative access of substrates to the enzyme region by a novel “two-dimensional” membrane design (8, 9), which permits both radial and axial diffusion of oxygen but only axial diffusion of glucose into the immobilized enzyme gel. These features allow the sensor to respond to glucose over a clinically useful concentration range even at very low tissue oxygen concentrations.

Variable microvascular perfusion

The signals of individual electrodes are affected by convection and diffusion of the substrates, in addition to their concentrations in blood. In tissues, glucose and oxygen are conveyed to the implant site by blood that perfuses the regional microvasculature, and then diffuse from capillaries to the sensor. Physiologic variations in blood flow associated with exercise, sleep, movement, hydrostatic changes, and local temperature changes affect the oxygen flux to both the glucose and the oxygen reference electrodes simultaneously, as does tissue remodeling, but the signal artifacts associated with these common physiological events are largely subtracted by our differential oxygen detector design. There was limited opportunity to test a range of physiologic variables in the present study, but the capability for artifact subtraction by the sensor can be further assessed in outpatient clinical applications.

Regardless of their respective concentrations in blood, glucose and oxygen are distributed heterogeneously in tissues at the microscopic level (11, 30). As shown in Fig. 5, there is a broad range of oxygen sensor signals between the 10th and the 90th percentile limits. Although oxygen sensors are uniform in fabrication and produce near-identical signals in vitro, when implanted, the sensors produce a range of signal values because of the specific microvascular pattern in the immediate neighborhood of each sensor (21). The array of paired glucose and oxygen sensors in the implant potentially provides a means for averaging these local spatial distributions of the substrates, although this feature was not used in the present study.

Sensor dynamic response

A key question is whether an implanted sensor can respond fast enough to follow physiologic blood glucose changes. For this sensor, in which the response of the sensor per se is rapid relative to glucose mass transfer within the tissues, the overall rate of response depends on achieving minimal tissue encapsulation and maintaining adequate tissue permeability.

To establish dynamic performance requirements of the sensor, we analyzed archival blood glucose data from stable and unstable diabetic subjects, as well as several nondiabetic controls (31, 32). These data consisted of blood glucose values obtained by regular venous blood collection every 3 to 5 min for 24- to 48-hour periods (33). The goal was to identify the most rapid excursions expected in diabetic subjects and define the characteristics that would be required of a continuously operated sensor to accurately reproduce these excursions. For glucose monitoring based on discrete blood sampling, it was determined that regular sampling at most every 12 to 15 min is necessary and sufficient to accurately reconstruct the most rapid physiologic blood glucose excursions, according to the classic Shannon-Nyquist sampling criterion [two regularly spaced samples per cycle of the most rapid frequency component (34)]. This suggests that a continuously operated sensor having a maximal 12- to 15-min delay should be equally effective in capturing blood glucose excursions. The average 11.8-min rising and 6.5-min falling delays seen with this sensor are well within this criterion, indicating that, on average, this system is capable of following all rapid blood glucose excursions expected in the minority of rapid responding diabetic subjects and, by extension, more typical slower excursions in most diabetic subjects.

Results from certain short-term subcutaneous sensor studies in humans have shown comparable delay values. Measurements in some studies have reported delay values from 4 to 10 min in humans (19), and an approach based on retrospective correlation between sensor signals and fingerstick assay values estimated a statistical delay of 12.5 min (20).

We have also shown that autoregressive moving-average methods based on previous blood glucose measurements can predict blood glucose values ahead of real time by as much as 20 min with quantifiable accuracy (35). This strategy could be used to further mitigate the effects of delay, if necessary in later applications.

For potential use with an artificial pancreas, the sensor described here is relatively rapid relative to the other components of a closed-loop system, namely, the serial processes of insulin delivery from a pump to a tissue site, insulin adsorption into blood (which, by itself, may be slow relative to the sensor), circulation of insulin and glucose to the peripheral tissues, and activation of the blood glucose change. Therefore, the sensor can have a key role in use of the artificial pancreas to counter hyperglycemic excursions. The convenience of a long-term, fully implanted sensor may also make an artificial pancreas more acceptable to a larger group of people with diabetes. Further, the ability of the sensor to detect and warn of hypoglycemia in a timely fashion will potentially increase the safety of automatic blood glucose control systems.

Conclusions

We have shown that, with appropriate design, an implanted glucose sensor can potentially operate effectively for long periods in the body. These experimental results and the understanding of the sensor function derived from animal studies provide a foundation for translation to human clinical investigation.

Materials and Methods

Device implantation

Individual sensor telemetry units were implanted in subcutaneous tissue sites in 20-kg anesthetized Yucatan minipigs by making an incision 5 cm long and 0.5 to 1 cm deep, retracting the skin, and exposing the dermal layers. A pocket was created between the subdermal fat and underlying muscle with blunt dissection while not disturbing the fascia. The implants were placed in this pocket with the sensor surface facing inward toward the muscle layer. Small polyester velour pads were previously fixed to the implant surface to encourage tissue ingrowth and attachment. Once the implant was seated in the pocket, the incision was sutured and the animal was wrapped with a protective bandage. A modified dual-lumen Hickman catheter (Bard Access Systems) was introduced into the central vena cava for blood sampling and fluid infusion, with the catheter ports exteriorized at the midscapular region. The catheter was maintained patent between uses with a dilute solution of heparin. Sterile technique was used in the procedures, and the National Institutes of Health Guide for the Care and Use of Laboratory Animals was followed for all animal activities.

Glucose excursion tests

Glucose excursion tests were performed once or twice weekly. Glucose excursions were achieved by IVGTTs administered by controlled central venous infusion of a 50% glucose solution. In nondiabetic animals, a concomitant infusion of a somatostatin analog (Bachem) was used to partially suppress endogenous insulin production. The results were blood glucose excursions that included a rapid rise from the normoglycemic value (~70 mg/dl) to a plateau of ~250 mg/dl, a dwell at the plateau value for ~20 min, and then a rapid unaided fall to the baseline because of the action of endogenous insulin, with occasional mild hypoglycemic undershoot. In diabetic animals, an intravenous insulin bolus was typically used to acutely drop blood levels from starting values of between 200 and 250 mg/dl to nadirs of between 50 and 100 mg/dl, in addition to IVGTT. Central venous blood was sampled every 5 to 10 min during excursions, and central venous plasma glucose values were determined with a Yellow Springs Instrument Company (YSI) 2300 STAT Plus glucose analyzer.

Conversion to diabetes

Pigs were made diabetic by infusion of streptozotocin (85 mg/kg) (Axxora). After conversion, animals were maintained on multiple daily subcutaneous and intravenous injections of insulin at typically 0.3 to 0.7 U/kg per day. Blood glucose samples were obtained frequently during the first day after induction of diabetes to assure avoidance of severe hypoglycemia, and multiple times daily thereafter.

Sensor calibration

As an optimum calibration interval was not known a priori, a protocol based on a fixed 10-day interval was used with glucose excursions for sensor calibration. The least-squares error between the sensor output and YSI assays of central venous plasma samples was used to determine the values of k1 and k2, and the resulting calibration was then used for the following 10 days. Correlation data obtained during a sensor response to a “calibration excursion” were not included in the accuracy determinations. The statistical analyses included only data collected during the sensor responses on days subsequent to calibrations.

Tissue visualization

Staining of conjugated antibodies used a labeled streptavidin-biotin immunoenzymatic antigen detection system (36, 37). The stained slides were imaged with an Olympus VANOX-S slide microscope at ×40 magnification with an attached Olympus E-30 DSLR with microscope lens mount. The resulting scale factor for the images was 0.34 μm per pixel. Constant exposure and sensitivity settings were maintained during the imaging to maintain identical light levels as indicated by the camera internal light meter. Once a slide was digitally imaged, five regions of 250-μm width and 500-μm depth adjacent to the sensor surface were selected for quantitative histology analysis.

The region visualized was selected at random, but regions with staining artifacts, tissue damage, or edge artifacts were avoided. Each image used a combination of color and morphology restrictions to identify stained pixels. These pixels were then divided by the total number of pixels in the image, and a quantitative value was obtained.

Footnotes

  • Citation: D. A. Gough, L. S. Kumosa, T. L. Routh, J. T. Lin, J. Y. Lucisano, Function of an implanted tissue glucose sensor for more than 1 year in animals. Sci. Transl. Med. 2, 42ra53 (2010).

References and Notes

  1. Funding: This work was supported by grants DK64570 and DK77101 [University of California San Diego (UCSD)] and DK77254 (GlySens) from the NIH. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institute of Diabetes and Digestive and Kidney Diseases or the NIH. Early research was supported by grants from the Juvenile Diabetes Research Foundation. Author contributions: Sensor design was done by D.A.G., J.Y.L., J.T.L., and T.L.R.; animal studies were conducted by J.T.L., T.L.R., and J.Y.L.; histology samples were processed and images were prepared by L.S.K.; device fabrication was carried out by T.L.R., J.T.L., and J.Y.L.; all authors participated in data analysis and manuscript preparation. Competing interests: D.A.G. is founder and consultant to GlySens Incorporated, an arrangement approved by UCSD in accordance with its conflict of interest policies. He is a co-inventor of patents related to the device reported here. J.Y.L. is the president and CEO of GlySens Incorporated and a co-inventor of patents related to the device. D.A.G., J.T.L., J.Y.L., and T.L.R. have equity interests in GlySens. GlySens is the owner and developer of the device and licensee of intellectual property for some aspects of the technology from UCSD. UCSD holds three patents related to this work.